SITRANS P DS III/P410 (7MF4.33.. 7MF4.34.. 7MF4.35..)
38
A5E03434626-04, 05/2015
gauge pressure input, with flush mounted diaphragm
Lower measuring limit
•
Measuring cell with silicone oil filling
100 mbar a/10 kPa a/1.45 psi a
•
Measuring cell with inert liquid
100 mbar a/10 kPa a/1.45 psi a
•
Measuring cell with neobee
100 mbar a/10 kPa a/1.45 psi a
Upper measuring limit
100% of maximum span
1)
Order the nominal measuring range with the order option Y01 for PROFIBUS PA or FOUNDATION Fieldbus.
Absolute pressure input, with flush-mounted diaphragm
Measured variable
Absolute pressure
Span (continuously adjustable) or
measuring range, max. operating pressure
and max. test pressure
Span
1)
Maximum operating
pressure MAWP (PS)
Maximum test pressure
43 … 1300 mbar a
4.3 ... 130 kPa a
17 … 525 inH
2
O a
2.6 bar a
260 kPa a
37.7 psi a
10 bar a
1 MPa a
145 psi a
160 … 5000 mbar a
16 ... 500 kPa a
2.32 … 72.5 psi a
10 bar a
1 MPa a
145 psi a
30 bar a
3 MPa a
435 psi a
1 … 30 bar a
0.1 ... 3 MPa a
14.5 … 435 psi a
45 bar a
4.5 MPa
653 psi a
100 bar a
10 MPa a
1450 psi a
Depending on the process connection, the span may differ from these values
Lower measuring limit
0 mbar a/kPa a/psi a
Upper measuring limit
100% of maximum span
1)
Order the nominal measuring range with the order option Y01 for PROFIBUS PA or FOUNDATION Fieldbus.
Input pressure transmitter with PMC connection
Measured variable
Gauge pressure
Span (continuously adjustable) or
measuring range, max. operating pressure
and max. test pressure
Span
1)
2)
Maximum operating
pressure MAWP (PS)
Maximum test pressure
0.01 … 1 bar
1 ... 100 kPa
0.15 … 14.5 psi
4 bar
400 kPa
58 psi
6 bar
600 kPa
87 psi
0.04 … 4 bar
4 ... 400 kPa
0.58 … 58 psi
7 bar
700 kPa
102 psi
10 bar
1 MPa
145 psi
0.16 … 16 bar
0.016... 1.6 MPa
2.3 … 232 psi
21 bar
2.1 MPa
305 psi
32 bar
3.2 MPa
464 psi
Lower measuring limit
•
Measuring cell with silicone oil filling
2)
100 mbar a/10 kPa a/1.45 psi a
•
Measuring cell with inert liquid
2)
100 mbar a/10 kPa a/1.45 psi a
•
Measuring cell with neobee
2)
100 mbar a/10 kPa a/1.45 psi a
Upper measuring limit
100% of maximum span
1)
Order the nominal measuring range with the order option Y01 for PROFIBUS PA or FOUNDATION Fieldbus.
2)
For PMC-Style Minibolt, the span should not be less than 500 mbar