71
PBD-51041059
SITRANS LG260 - Operating Instructions
44284-EN-171021
Protection rating
Ʋ
unassembled
IP 20
Ʋ
mounted in the housing without lid
IP 40
Materials
Ʋ
Housing
ABS
Ʋ
Inspection window
Polyester foil
Functional safety
SIL non-reactive
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone, factory setting
CET
Max. rate deviation
10.5 min/year
Additional output parameter - Electronics temperature
Output of the values
Ʋ
Indication
Via the display and adjustment module
Ʋ
Analogue
Via the current output
Ʋ
Digital
Via the digital output signal (depending on the electron-
ics version)
Range
-40 … +85 °C (-40 … +185 °F)
Resolution
< 0.1 K
Accuracy
±3 K
Voltage supply
Operating voltage
8 … 30 V DC
Power consumption
< 500 mW
Reverse voltage protection
Integrated
Potential connections and electrical separating measures in the instrument
Electronics
Not non-floating
Ground terminal
Galvanically connected with the metal process fitting
Galvanic separation between electronics and metal housing parts
Ʋ
Reference voltage
500 V AC
Electrical protective measures
Protection, depending on housing version
Ʋ
Plastic housing
IEC 60529 IP 66/IP 67 (NEMA Type 4X)
Ʋ
Aluminium housing; stainless steel
housing - Precision casting
IEC 60529 IP 66/IP 68 (0.2 bar), NEMA Type 6P
Connection of the feeding power supply
unit to networks of overvoltage category
III