OPERATION
6
87
SITRANS FX330
www.siemens.com/flow
05/2016 - A5E36751269-AA
Status
type
Event group
Single event
Description
Actions to eliminate the
event
F
Sensor
Fatal Sensor Error
An internal error in the
sensor module has occured.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
No Temperature Sensor
The internal temperature
sensor (PT1000) is not
connected or its resistance
is too high. The process
temperature cannot be
measured.
Check the pick-up and its
connection. Contact the
manufacturer, if necessary.
Sensor Comm. Error
Error in the internal bus
communication or hardware
failure.
For remote version: Check
the sensor connection.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
No/Invalid Pressure Sensor
Pressure sensor not
available or invalid
communication
Check the connection of the
pressure sensor. Perform a
hard reset. If the message
recurs, contact the
manufacturer.
Corrupt Sensor Parameter
Flow sensor parameters are
inconsistent
Check flow sensor
parameters.
Temperature Sensor Error
The resistance of the
internal temperature sensor
(PT1000) is too low.
Check the pick-up and its
connection. Perform a hard
reset. Contact the
manufacturer, if necessary.
No Measurement Value
No connection to flow
sensor
Contact the manufacturer.
F
Electronics
Fatal Converter Error
An electronics or hardware
failure has occured.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
CO DM Error
Internal electronic failure
CO DDM Error
Internal electronic failure
CO Unknown Error
Internal electronic failure
CO ADC Error
Internal electronic failure
Internal Comm. Error
Error in the internal bus
communication.
Inconsistent Parameters
Inconsistent data in the
parameter memory after
editing via HART
®
interface
Check the parameters and
make sure that the set of
data is consistent. Additional
information on the error can
be called via HART
®
interface.
Sensor Comm. Error
Error in the internal bus
communication or hardware
failure
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Power Supply Error
Internal voltage too low to
switch on sensor power
supply
Check power supply
connection or replace the
electronics.
Sensor/Conv. FW Mismatch
Incompatible firmware
versions in the sensor and
the converter module
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Max. Number of Restarts
The device could not start
the measuring mode after a
cold starting.
Perform a hard reset once
again. If the message
recurs, contact the
manufacturer.