3-15
Section 3
CQO:1010PVXFM-3
3.3.9 USING 1012T MOUNTING TRACKS
The 1012TP and 1012THP Mounting Tracks provide a rigid mounting platform for Series 1011 universal
or high precision size A or B transducers. The mounting tracks service pipe sizes up to a maximum of
5.00” (140 mm) outer diameter. Operating temperatures are supported up to 250°F (121°C). The as-
sembly consists of lightweight aluminum track rails with integral transducer clamping screws. Attached
index pins enable positive locating of the transducers at fixed spacing locations. Roller-chains and ten-
sion screws secure the assembly to the pipe. The following instructions refer to 1011 universal trans-
ducers. Please refer to Engineering Drawing 1012THP-7A for reflect mounting of high precision trans-
ducers and 1012THP-7B for direct mounting of high precision transducers.
The 1012T mounting tracks support both Direct and Reflect mounting modes. The flow computer recom-
mends the appropriate transducers, mounting track and mounting mode, based on the pipe data en-
tries. Refer to the instructions in paragraph 6.4 for details on the Transducer Installation procedure. If
necessary, review paragraphs 3.3.2 through 3.3.4 for details on how to select and prepare a mounting
location on your pipe.
Installing a 1012T Mounting Track in Reflect Mode
Paragraph 3.4 describes the Transducer Installation procedures that lead up to the automatic selection of
transducers, mounting mode and spacing method. Examine the figure below, which shows a typical
Pick/Install Xdcr menu screen. Note the automatic assignment of model numbers for the transducer and
mounting track, plus the designation of the number index
.
This example requires a Model 1012T Mounting Track to accommodate size B3 universal transducers.
Note the reported number index. You will be inserting an index pin into this hole on the track rail to
position one of the transducers (see diagram on next page).
1. Perform all required menu steps up until the point where you respond to the [Install Completed?]
prompt. Note the reported number index. You will be inserting index pins into this hole and the
reference hole on the track rail (see diagram on next page). Check to ensure that you have a matched
set of transducers. They both should have the same S/N number but marked with either an “A” or “B”
(e.g., 100A and 100B).
Note selection of
type, size and
mounting mode.
Note selection of
mounting track part
number and number
index.
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Transducer Model
1011 Universal
Transducer Size
B3
Xdcr Mount Mode
Reflect
Spacing Offset
Minimum
Number Index
6
Spacing Method
Track 1012TP
Ltn Value (in)
0.581
Install Completed?
No
Empty Pipe Set
Channel Not Setup
Zero Flow Adjust
Channel Not Setup
Содержание SITRANS FUH1010PVDX
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 12: ......
Страница 20: ......
Страница 24: ......
Страница 32: ......
Страница 184: ......
Страница 204: ......
Страница 206: ......
Страница 208: ......
Страница 216: ......
Страница 219: ...21614 C 1010X 7 CAUTION ...
Страница 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Страница 223: ...21614 C 1010DX 7 CAUTION ...
Страница 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 234: ...1011HNFS 7 21614 C ...
Страница 235: ...1011HNFS 7 C 21614 ...
Страница 237: ......
Страница 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Страница 242: ...1011NFPS 8 21614 C ...
Страница 243: ......
Страница 245: ...21614 C 1012 57 8 CHANNEL 1 UP DN UP 1012CNF TEMP SENSOR HOUSING CABLE JUNCTION BOX OUTLINE DIMENSIONS ...
Страница 247: ...1012FN 8 OUTLINE DIMENSIONS 1011 SERIES DEDICATED DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 248: ...1012FNH 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE MOUNTING FRAME PART NUMBER L H W Loa X 21614 C ...
Страница 249: ......
Страница 250: ......
Страница 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Страница 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Страница 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Страница 258: ...1012F DB 7 21614 C ...
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Страница 263: ......
Страница 264: ......
Страница 265: ......
Страница 266: ......
Страница 267: ......
Страница 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Страница 270: ......
Страница 271: ......
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ......
Страница 286: ......
Страница 304: ......
Страница 307: ......