6-14
CQO:1010PVXFM-3
Section 6
To clear ALL Dynamic Memory using the Magnetic Wand:
CAUTION: Before proceeding further it is essential to understand that this function eliminates
ALL data stored in RAM. This means that all saved Site Setups including the site data
of a flow-calibrated site will be erased! In addition, the entire Datalogger file plus
any custom factory or user-created pipe or transducer tables will be eliminated. The
impact of this is such that we strongly recommend that you consult the Technical
Service Department before continuing with this procedure. If you choose to con-
tinue, be aware that you will have to create a new Site Setup, re-enter all site specific
parameters including pipe or transducer tables, plus all desired Meter Facilities en-
tries.
1. Turn off power (if it is currently on) and then reapply power to the system.
2. Perform steps 1 and 2 in the “Clear Active Memory” instructions above.
3. Trigger . Note that the greater than (>) symbol changes to a (:) colon. Trigger
.
The display
switches to [Clr Saved Data?] as shown on the left-hand figure (see below).
3. Trigger . Note that the greater than (>) symbol changes to a (:) colon. Trigger to change
the second line to [Yes] (as shown in the right-hand figure above).
4. Trigger
to clear the contents of Active Memory (but preserve any stored Site Setups).
5. To restore operation, create a new site setup or recall a stored site setup and then re-select
any Meter Facilities items (e.g., RS-232 setup parameters) that you may have edited origi-
nally.
Clr Active Memory?
:Yes
Clr Active Memory?
>No
Clr Saved Data?
>No
Clr Saved Data?
:Yes
Содержание SITRANS FUH1010PVDX
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 12: ......
Страница 20: ......
Страница 24: ......
Страница 32: ......
Страница 184: ......
Страница 204: ......
Страница 206: ......
Страница 208: ......
Страница 216: ......
Страница 219: ...21614 C 1010X 7 CAUTION ...
Страница 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Страница 223: ...21614 C 1010DX 7 CAUTION ...
Страница 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 234: ...1011HNFS 7 21614 C ...
Страница 235: ...1011HNFS 7 C 21614 ...
Страница 237: ......
Страница 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Страница 242: ...1011NFPS 8 21614 C ...
Страница 243: ......
Страница 245: ...21614 C 1012 57 8 CHANNEL 1 UP DN UP 1012CNF TEMP SENSOR HOUSING CABLE JUNCTION BOX OUTLINE DIMENSIONS ...
Страница 247: ...1012FN 8 OUTLINE DIMENSIONS 1011 SERIES DEDICATED DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 248: ...1012FNH 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE MOUNTING FRAME PART NUMBER L H W Loa X 21614 C ...
Страница 249: ......
Страница 250: ......
Страница 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Страница 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Страница 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Страница 258: ...1012F DB 7 21614 C ...
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Страница 263: ......
Страница 264: ......
Страница 265: ......
Страница 266: ......
Страница 267: ......
Страница 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Страница 270: ......
Страница 271: ......
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ......
Страница 286: ......
Страница 304: ......
Страница 307: ......