6-17
CQO:1010PVXFM-3
Section 6
z
Access the [Pick/Install Xdcr] menu. Check the [Transducer Size] menu cell. If necessary, enter the
[Xdcr Size] option list and pick the transducer under test.
z
Check the [Xdcr Mount Mode] menu cell. Adjust to match the simulator chart above.
z
Move the highlight down to [Install Complete?] by triggering the <Down Arrow>. Access the option
list. Move the cursor to [Install] by triggering the <Up or Down Arrow>. Trigger the ENTER sensor to
start the transducer install routine.
NOTE: Since Sizes A and C transducers are installed in Reflect Mode, you will see a pop-up
window that prompts you to: [Use Actual Zero]. You can ignore this by triggering the
<Down Arrow>.
Upon the completion of the [Install] procedure, the computer should report a sonic velocity within the
range of approximately 1350m/s to 1700 m/s (depending on the transducer size under test). Next,
check the
[Diagnostic Data]
menu. The [Valc %] item must be >35 for dependable operation. In addi-
tion, note the reading, then compare it to the [Vf max] item in the
[Diagnostic Data/Site Setup Data]
menu. The value of the reading should be less that 2% of the published Vf max. Confirming these values
certifies that the entire system (computer, transducers, cables) is operating correctly. The investigation
should proceed to a review of all site conditions to locate the operating problem.
6.9.2 USING THE 996PSP PIPE SIMULATOR
Using the proper coupling compound, mount the transducers on the pipe simulator as shown below.
Slide each transducer until it presses against a pin-stop. Use the clamping screws to hold the transduc-
ers in place.
Part
Xdcr Pipe OD
Pipe
Wall
Mount
Spacing
Number
Number
Size
(in)
Mat’l
Thk (in) Mode
Offset
Index
1012TB-1
A
0.650
Steel*
0.100*
Reflect
Nominal
7
1012TB-1
B
1.150
Steel*
0.100*
Direct
Minimum
4
1012TB-2
C
2.000
Steel*
0.100*
Reflect
Nominal
11
1012TB-2
D
3.500
Steel*
0.100*
Direct
Minimum
Use Ltn
1012TB-1 & 2 Universal Transducer Test Block Chart (English)
*System Default
1012TB-1
A
16.5
Steel*
2.54*
Reflect
Nominal
7
1012TB-1
B
29.2
Steel*
2.54*
Direct
Minimum
4
1012TB-2
C
50.8
Steel*
2.54*
Reflect
Nominal
11
1012TB-2
D
88.9
Steel*
2.54*
Direct
Minimum
Use Ltn
1012TB-1 & 2 Universal Transducer Test Block Chart (Metric)
Part
Xdcr Pipe OD
Pipe
Wall
Mount
Spacing
Number
Number
Size (mm)
Mat’l
Thk (mm)
Mode
Offset
Index
*System Defaults
Содержание SITRANS FUH1010PVDX
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 12: ......
Страница 20: ......
Страница 24: ......
Страница 32: ......
Страница 184: ......
Страница 204: ......
Страница 206: ......
Страница 208: ......
Страница 216: ......
Страница 219: ...21614 C 1010X 7 CAUTION ...
Страница 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Страница 223: ...21614 C 1010DX 7 CAUTION ...
Страница 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 234: ...1011HNFS 7 21614 C ...
Страница 235: ...1011HNFS 7 C 21614 ...
Страница 237: ......
Страница 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Страница 242: ...1011NFPS 8 21614 C ...
Страница 243: ......
Страница 245: ...21614 C 1012 57 8 CHANNEL 1 UP DN UP 1012CNF TEMP SENSOR HOUSING CABLE JUNCTION BOX OUTLINE DIMENSIONS ...
Страница 247: ...1012FN 8 OUTLINE DIMENSIONS 1011 SERIES DEDICATED DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 248: ...1012FNH 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE MOUNTING FRAME PART NUMBER L H W Loa X 21614 C ...
Страница 249: ......
Страница 250: ......
Страница 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Страница 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Страница 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Страница 258: ...1012F DB 7 21614 C ...
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Страница 263: ......
Страница 264: ......
Страница 265: ......
Страница 266: ......
Страница 267: ......
Страница 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Страница 270: ......
Страница 271: ......
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ......
Страница 286: ......
Страница 304: ......
Страница 307: ......