Because of this small, but increasing contribution to F
0
once T reaches 100 ºC, and the fact
that in this trace the temperature exceeds 121.11ºC, the F
0
value reaches 15 at time t = 20
minutes, at which point the sterilisation is complete, and cooling begins.
For some applications it is allowed to report the F
0
value that is finally achieved as
the sample cools back to the start temperature, in this example 100 ºC, and so the
contribution under the curve up to t = 22 minutes would be included in the reported
F
0
. The final F
0
value might then be 16.1, say, with 15 having been achieved up to t =
20 minutes, and the extra 1.1 having accumulated during the cooling back to 100 ºC
between t = 20 minutes and t = 22 minutes.
The terms used in setting up the F
0
totaliser, together with their default values, are summa-
rised in this table.
For setting up the Fo calculation refer to the
Term
Definition
Default
values
Allowed
Range
F
0
The equivalent time in minutes at 121.11ºC that would pro-
duce the same degree of sterilisation. This is the totaliser out-
put value
z factor
The temperature increase in ºC above the reference tempera-
ture that will produce a 10-fold increase in the rate of sterilisa-
tion
10
5 - 20
Start
temperature
The temperature in ºC above which the contributions to F0
can be counted
100
90 - 120
Reference
Temperature
The reference temperature for sterilisation, normally
121.11
ºC, can be altered if required.
121.11
ºC
Include
cooling phase
If TRUE, contributions to F
0
can be included after the target
value has been reached and while the temperature is still
above the start temperature. If FALSE, these contributions are
not included
FALSE
-
Completion
Value
The maximum
F
0
totaliser value can be set if “Include
cooling phase” is not selected, the totaliser will stop at
this value of F
0
.
15
1 - 999
Содержание SIREC D200
Страница 2: ......
Страница 176: ...170 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Firmware Credit System ...
Страница 213: ...43 TV 25 35 GLO Iss 4 Dec 06 UK 207 A5E01001767 04 SIREC D400 Recorder 14 15 32 46 50 30 31 35 36 28 55 ...
Страница 214: ...208 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 52 11 23 33 39 5 6 RS485 USB host Ethernet 1 2 3 4 ...
Страница 218: ...212 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 SIREC D200 Recorder 16 5 3 1 4 2 24 49 9 45 19 6 49 ...
Страница 222: ...216 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 238: ...232 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 258: ...252 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 270: ...264 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 274: ...268 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Modbus Function Codes ...
Страница 284: ...278 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 285: ......