27
MAN-100472
RD150 - Operating Instructions
57030-EN-180717
After a reset, certain parameter adjustments made by the user are
reset.
The following reset functions are available:
Delivery status:
Restores the parameter settings at the time of ship-
ment Ex factory including the order-specific settings.
Basic settings:
Resetting the parameter settings to the default val-
ues of the respective instrument.
The following table shows the default values of the instrument. De-
pending on the instrument version or application, all menu items may
not be available or some may be differently assigned:
Reset - Setup
Menu item
Parameter
Default value
Measurement
loop name
Display
Display
Language
English
Order-specific
Displayed value
Signal current
Backlight
Switched off
Damping
Integration time
0 s
Scaling
Scaling size
%
Scaling format
20 mA correspond to 100.00 %
4 mA correspond to 0.00 %
Lock adjustment
Released
Reset - Additional settings
Menu item
Parameter
Default value
HART
HART mode
Primary Master
HART address
Address 0
With this function, the display settings are copied.
The following parameters or settings are saved:
•
All parameters of the menu "
Setup
" as well as the menu item
"
Additional adjustments - HART mode
"
Additional settings -
Reset
Additional adjustments -
Copy display settings