Replacing important components
Replacing the collimator
5
In the case of damage, the collimator has to be replaced completely.
Proceed as described in the following:
•
Loosen two screws (5/Fig. 3) and remove the rear panel.
•
Mark the 5 connecting cables of the collimator and disconnect them.
•
Remove the two screws on both sides (6/Fig. 3) of the upper cover.
•
Turn the two knobs (8/Fig. 4) in their final position and remove them.
•
Remove the front panel after having loosened the two screws (7/Fig. 4).
Caution:
Front panel is connected to collimator via cables (9/Fig. 5).
•
Remove the collimator after loosening the four Allen screws (arrows /Fig. 2/Fig. 3).
•
Attach the new collimator, fasten it with the four screws (arrows /Fig. 2/Fig. 3) and center
it.
•
Connect the connecting cables and screw down the rear panel.
•
Check agreement of light and radiation field and make adjustment if required
(see “Agreement of light and radiation field”).
Hold the collimator.
Lift upper cover and pull out connecting cable through the case.