
POLYMOBIL III
RXB8-115.061.02
Page 24 of 24
Siemens AG
Rev. 01
10.95
CS PS 24
Medical Solutions
4 - 24
Notes on trouble-shooting
Checking the control elements on the collimator
4
•
Check the light-beam localizer operated both at the control panel and at the collimator.
To do so, press the
button (6/Fig. 10 or 7/Fig. 9),
the light-beam localizer lamp lights up for 20 s
±
1s.
•
Check if the double-slot diaphragm can be rotated
±
45°.
•
Check that format adjustment (blades) is easy and smooth by turning the collimator
adjustment knobs (8/Fig. 9).
6
Fig. 9
Fig. 10
7
8