Notes on trouble-shooting
4 - 23
Siemens AG
RXB8-115.061.02
Page 23 of 24
POLYMOBIL III
Medical Solutions
Rev. 01
10.95
CS PS 24
Checking and readjusting the counterweight
4
•
Unlock the support arm (1/Fig. 7) and release the turning knob (2/Fig. 7).
Without any accessories attached, the support arm should be easy to move across the
entire movement range and stop in any position.
Readjusting the counterweight with accessories attached
4
If the support arm does not stop in any position with accessories attached, readjust the
spring tension for the counterweight.
•
Take the cover (3/Fig. 7) off the stand.
•
Completely release the additional brake on the support arm (2/Fig. 7).
•
Move the support arm into the horizontal position.
•
Tighten the screw by means of a 17mm fork wrench (4/Fig. 8).
Adjust the spring tension so that the support arm is balanced in the horizontal position,
i.e. the forces needed for lifting and lowering the tube unit are equal.
4
1
2
3
Fig. 7
Fig. 8