
POLYMOBIL III
RXB8-115.061.02
Page 6 of 24
Siemens AG
Rev. 01
10.95
CS PS 24
Medical Solutions
4 - 6
Notes on trouble-shooting
Checking the supply voltages
4
•
When the system is switched on, relays HR and K1 on D920 pull in and supply
transformer T1, power supply M6 and the intermediate circuit with line voltage.
•
Measure the 24 V supply voltage with the digital multimeter on D920 at MP "+ 24 V" and
"0V"..
•
Measure the filament circuit voltage supply on the electrolyte capacitor
C46 at +/ - on D920.
Measured voltage
Measured voltage approx. 28.2 V
Tolerance
±
5 V
in stand-by mode approx. 18.5 V
with preparation approx. 16.4 V
Tolerance
±
4 V
Tolerance
±
3 V
D920 (X038I-92D)