
POLYMOBIL III
RXB8-115.061.02
Page 10 of 24
Siemens AG
Rev. 01
10.95
CS PS 24
Medical Solutions
4 - 10
Notes on trouble-shooting
Oscillating current measurement
4
•
POLYMOBIL
OFF
•
Connect oscilloscope on D920 to MP "I
s
" and "GND".
•
POLYMOBIL
ON
•
Trigger exposure with default values 60 kV, 10 mAs.
Main inverter ok
•
During the kV rise
•
After the kV rise
Integrated current transformer.
1V 50 A
NOTICE
1 V 50 A
1 V 50 A