Notes on trouble-shooting
4 - 3
Siemens AG
RXB8-115.061.02
Page 3 of 24
POLYMOBIL III
Medical Solutions
Rev. 01
10.95
CS PS 24
•
Checking the fuses on D920
Loosen the 4 Allen screws (2/Fig.1, 4 mm key size) and remove the cover.
Check the following fuses:
F1
2 AT
Closing circuit
F2
2 AT
Closing circuit
F3
0.25 AT
T1 secondary
F5
4 AT
Filament circuit
F6
3.2 AT
24 V / Main inverter control
F7
10 AT
Light-beam localizerr
D920 (X038I-92D)