MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 4 of 10
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
11 - 4
Startup Protocol
Startup Settings
11
Tube high voltage, current and mAs values
11
Results from the chapter "Tube high voltage, current and mAs values" on Page 6 - 1.
kV
mAs
Anode/
filter
Measured values
Result
kV must be
&
5%
mAs must be
&
10%
30 kV
20 mAs
Mo/Mo
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
30 kV
20 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
35 kV
100 mAs
Mo/Mo
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
35 kV
100 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
25 kV
100 mAs,
Mo/Mo
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
25 kV
100 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
30 kV
10 mAs
Mo/Mo
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
Tab. 2
Tube high voltage, current and mAs values