Mechanical Checks
5 - 5
Siemens AG
SPB7-250.815.02
Page 5 of 12
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
Mechanical Checks
5
Rotary motion
5
The rotary motion with the detector wing has to be checked.
1.
System
OFF
.
The rotary motion and the vertical adjustment of the swivel arm system will be
blocked.
2.
System
ON
, detector table installed and projection angle set to minimum with
3.
Check the rotary motion of the swivel arm system by pressing buttons (2/Fig. 6)
one by one.
The rotation of the swivel arm system will stop when the button is released or
when the end position is reached (
+135 degrees clockwise
and -
180 degrees
counterclockwise
). The projection angle is shown on the display at the lower part
of the stand.
Fig. 6
Swivel arm system
2
1
2
3
4
3