MAMMOMAT stand.
The Service PC is required for MAMMOMAT stand service software.
A description of the various service software packages can be found in the MAMMOMAT
Novation
DR
Software
manual (
SPB7-250.816.02...
).
Delay times between two exposures
2
The delay times listed below must be followed to prevent the tube from overheating.
When the generator is switched off with the service PC con-
nected,
wait approximately 5 seconds before switching it on again
(The LED’s H3 and H4 have to be off).
Exposure mAs value
Delay time between two exposures (seconds)
max. 100
min. 15
max. 200
min. 30
max. 300
min. 45
max. 400
min. 60
max. 500
min. 75
Tab. 1
Delay times between two exposures
NOTE