MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 8 of 10
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
8 - 8
Checking dose settings
Adjust the H and D Settings
8
To adjust the
H
value, go to
Modify low dose scaling
(80, 70, 45) and
D
value, go to
Modify high dose scaling
(100, 80, 60)
and
enter the new values, separated by a comma, e.g.
100,80,60
. The values represent the
factor in % for (Mo/Mo, Mo/Rh, W/Rh
).
Three settings
try
(short-term, is valid for one measurement only),
save
(long-term, valid
also after reboot) and
load
(effective, valid until the system is rebooted) can be used in
the drop-down menu above the values.
To change the values permanently, enter the correct values and select
save
,
then use
load
to load them before you start another exposure.
Calculation of the new Mo/Mo dose scaling value:
NewScalingValue = OldScalingValue * DoseValue
Nominal
/ DoseValue
Measured
DoseValue
Nominal
for H = 130
µ
Gy
DoseValue
Nominal
for D = 190
µ
Gy
Calculation of the new Mo/Rh, W/Rh dose scaling values:
NewScalingValue = OldScalingValue * PixelValue
Mo/Mo
/ PixelValue
Measured
Proceed as follows:
1.
Go into the service software and select
Utilities
-> Escape to OS
and enter the
following command in the
Command
field.
start
explorer
http://brick
2.
Select
Brick -> AEC Configuration.
3.
The BRICK software switches to a higher user level, thus a Username and
Password have to be entered.
Username:
root
Password:
******