MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 2 of 6
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
6 - 2
X-ray tube checks
Oscilloscope diagram examples
Fig. 2
Fig. 3
30 kV
1 V/T (1 T <=> 5 kV)
30 kV
1 V/T (1 T <=> 5 kV)
130 mA
1 V/T (1 T <=> 40 mA)
166 mA
1 V/T (1 T <=> 40 mA)
160 ms
20 ms/T
128 ms
20 ms/T
Fig. 4
Fig. 5
35 kV
1 V/T (1 T <=> 5 kV)
35 kV
1 V/T (1 T <=> 5 kV)
112 mA
1 V/T (1 T <=> 40 mA)
132 mA
1 V/T (1 T <=> 40 mA)
0.9 s
0.2 s/T
0.72 s
0.2 s/T
30kV, 20mAs, Mo/Mo
30kV, 20mAs, W/Rh
35kV, 100mAs, Mo/Mo
35kV, 100mAs, W/Rh