MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 4 of 10
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
8 - 4
Checking dose settings
Checking the AEC dose settings
8
The dose values for the AEC settings (
H
&
D
selection on the control console) have to be
checked.
As a default from the factory, the following
BRICK / AEC Configuration
values are set
(for further details see the
Software Brick V...
manual SPB7-250.816.04...):
•
H = (80, 70, 45)
•
D = (100, 80, 60)
Checking the H & D dose settings for Mo/Mo.
8
Proceed as follows:
1.
Mount the Collimator Mounted Plexi phantom and the compression plate
simulator.
2.
Register or create a patient and submit with
Exam
.
Fig. 3
Calibration - Patient register