DCA Vantage Operator’s Guide
91
Qua
lit
y Cont
rol
Q
u
al
it
y C
o
nt
rol
To scan the reagent cartridge, refer to
Scanning the Reagent Cartridge
To insert the reagent cartridge, refer to
Inserting the Reagent Cartridge into
Figure 30 Scanning the Microalbumin/Creatinine Control Card
Quality Control Errors
If the control results fall outside the values stated in the package insert, the
following sources of error may have occurred:
1
Barcode track dot
Possible Cause
Corrective Action
Deterioration of the
reagent cartridge
test areas due to
exposure to light,
ambient moisture,
or heat.
Use a new reagent cartridge to repeat the quality control
procedure.
Use a fresh box of reagent cartridges, or a new lot.
If the new reagent cartridge fails to give results within the
expected values, proceed to the next possible cause.
Deterioration of the
control solution.
Use a fresh control solution to repeat the quality control
procedure.