DCA Vantage Operator’s Guide
129
Tr
oubles
hoot
ing
Troubleshooting Reagent Messages
These message display when an error occurs with the reagent cartridge.
Errors Codes and Corrective Actions
This section of the guide lists error codes and messages, along with a
description and corrective actions.
Touchscreen not
responding or not
responding
correctly.
Touchscreen
needs to be
calibrated.
Run the touchscreen calibration.
Refer to
Symptom
Possible Cause
Action
Reagent not
absorbed by
cartridge filter at
the end of the
test.
The
DCA Vantage
system is not on
a level work
surface.
Verify that the system is properly
installed on a level surface.
The filter in the
capillary holder is
not absorbent.
Contact your local technical support
provider.
Thermal Dot has
reacted (HbA
1c
reagents only)
Product was not
shipped or stored
properly.
There are 2 indicators: outside and
inside. The outer shipper has an
over temperature indicator. If this
indicator is red or black, the product
was not stored properly. Check the
inside indicator to confirm. If the
inside indicator is red, the
cartridges should be discarded. The
outside indicator can be red or
black, while the inside indicator is
normal. In this case, the cartridges
can be used.
Symptom
Possible Cause
Action
Содержание DCA Vantage
Страница 10: ...10 DCA Vantage Operator s Guide ...
Страница 24: ...24 DCA Vantage Operator s Guide ...
Страница 40: ...40 DCA Vantage Operator s Guide ...
Страница 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Страница 120: ...120 DCA Vantage Operator s Guide ...
Страница 150: ...150 DCA Vantage Operator s Guide File Management ...
Страница 166: ...166 DCA Vantage Operator s Guide ...
Страница 192: ...192 DCA Vantage Operator s Guide ...
Страница 210: ...210 DCA Vantage Operator s Guide Glossary ...