DCA Vantage Operator’s Guide
199
Computer Interface (LIS) C
o
mputer Interface
(LIS)
Appendix G: Computer Interface (LIS)
General Information
You can connect the DCA Vantage system to a host computer or laboratory
Information System (LIS). The DCA Vantage system supports a serial
connection or an Ethernet connection between instrument and host. The serial
interface is compatible with the EIA-232-D Specification. The Ethernet
interface is established through the instrument’s RJ45 internet connection.
Messages transferred from the DCA Vantage instrument to the host include:
•
HbA
1c
test result records
•
Microalbumin/Creatinine test result records
•
Filter Test Cartridge result records
•
Control Test result records
•
Cartridge Lot number
•
System serial number
Messages transferred from the host to the DCA Vantage system include:
•
Force Control remote commands
•
Remote Lockout remote commands
•
List of Operators
•
QC test request
Ethernet Connections Using POCT1-A
The DCA Vantage system uses the POCT1-A standard to link to laboratory
information systems. The POCT1-A standard automates transmission of
information including date and time, result, patient ID, and Operator ID, to
support a full electronic patient record. POCT1-A is the default Ethernet
setting.
Содержание DCA Vantage
Страница 10: ...10 DCA Vantage Operator s Guide ...
Страница 24: ...24 DCA Vantage Operator s Guide ...
Страница 40: ...40 DCA Vantage Operator s Guide ...
Страница 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Страница 120: ...120 DCA Vantage Operator s Guide ...
Страница 150: ...150 DCA Vantage Operator s Guide File Management ...
Страница 166: ...166 DCA Vantage Operator s Guide ...
Страница 192: ...192 DCA Vantage Operator s Guide ...
Страница 210: ...210 DCA Vantage Operator s Guide Glossary ...