5-26
5
Instrument Set Up
Authorized Operator
This section describes how to set up
use of operator IDs and add, edit, or
delete the list of operator IDs. When
enabled, the system permits only
allowed operators to perform patient
tests, QC tests (when using the
connector), recall results, or modify
system settings. Operators gain
access by entering their ID.
The Clinitek analyzer
stores 700 operators.
NOTE:
The Operator ID is never
printed or displayed with patient
results. If you wish to associate the
Operator’s Name with patient
results, enable Operator Name in
Custom Settings-Operator
screen
1 of 5.
CAUTION
Once the Operator ID and Operator
Name settings are made, do not
change the Operator ID setting.
If you change the Operator
ID setting, all patient results are
erased.
Setting Operator IDs
To set up operator IDs, perform the
following steps:
1. At the
Select Ready
screen,
touch
Instrument Set Up
.
The
Choose Settings
screen
displays.
Instrument Settings
2. Use the arrow keys to select
Instrument Settings
.
3. Touch
Select
.
The
Instrument Settings
screen
displays.
Содержание CLINITEK Status
Страница 1: ...Operator s Manual REF 10379682 ...
Страница 2: ......
Страница 3: ...134798 Rev A 2009 07 Operator s Manual ...
Страница 26: ...2 6 ...
Страница 28: ...3 2 ...
Страница 56: ...4 28 ...
Страница 94: ...5 38 ...
Страница 98: ...6 4 ...
Страница 108: ...7 10 ...
Страница 110: ...8 2 ...
Страница 132: ...10 14 10 Appendices Appendix E System Overview Principles ...
Страница 133: ...10 15 10 Appendices Appendix E System Overview Principles ...
Страница 134: ...10 16 10 Appendices Appendix E System Overview Principles ...
Страница 135: ...10 17 10 Appendices Appendix E System Overview Principles ...