4-20
4
Testing
Entering the Strip Lot Number
and Expiration Date
To enter strip lot information for a
second strip test, perform the
following steps:
1. At the
Select
screen, touch
Strip
Test
.
The
Strip
screen displays.
2. To use the last strip number and
begin the test, touch
Use Last
Lot
.
To enter new strip data, touch
Enter new lot and expiration
.
The
Strip Lot
screen displays.
3. Enter the strip lot number.
Use the alpha keyboard to enter
text.
To enter numeric text, touch
123
.
4. Select
Enter
.
The
Strip Expiration
screen
displays.
5. Use the arrow keys to indicate
the strip expiration date.
6. Touch
Enter
.
The
Prepare Test
screen
displays.
7. Touch
Start
.
See above.