Siemens AG
SPR2-330.815.01.03.02
ARCADIS Avantic
12.06
CS PS SP
Protocols
57
Page 57 of 66
Medical Solutions
Measuring circuit
0
Direct measurement
Fig. 23:
Measuring circuit for direct measurement of the system leakage current in compliance
with DIN VDE 0751-1/2001-10, Fig. C5 for protection class I.
Pos. 1
= System
Pos. 2
= Application part type B (if available)
Pos. 3
= Measurement setup (integrated into measuring device)
Fig. 24:
Measuring circuit for direct measurement of the system leakage current in compliance
with DIN VDE 0751-1/2001-10, Fig. C5 for protection class II.
Pos. 1
= System
Pos. 2
= Application part (if applicable)
Pos. 3
= Measurement setup (integrated into measuring device)