NOTE
With the introduction of the Hibernation Mode, system configura-
tion changes are not applied if the system is switched off and then
switched on again using the power buttons on the monitor cart.
To correctly apply the modified data, perform a system shutdown.
Open the <Options> - <End Session> menu and click the “Shut-
down system” button. Immediately thereafter, press the Power off
button on the monitor cart.
The system shuts down and switches off automatically.
NOTE
Once the configuration has been changed, the label affixed to the
footswitch has to be removed and replaced by the new label
included in the delivery volume. In the case of an "enhanced" foot-
switch, there are two labels on the pedals that have to be replaced.
•
Remove the label affixed to the footswitch. In the case of an "enhanced" footswitch,
remove both labels from the pedals.
•
If necessary, remove any remaining adhesive where the labels were removed.
•
Affix the new label to the footswitch. In the case of an "enhanced" footswitch, affix both
labels to the pedals.
•
If the footswitch configuration was changed, check the footswitch pedal function.
¹
The footswitch pedal functions have to correspond to the functions indicated on
the newly affixed label(s).
Country-specific adaptation of the maximum tube current
0
NOTE
A maximum tube current limit is required - as far as known - only
in Denmark.
•
Start service program and log in as described under "Service Login".