Notes on the warning signs
•
Before commissioning, the English warning label “LASER RADIATION – DO NOT
STARE INTO BEAM” is to be replaced with a warning label in a language that the
operators of the system understand. Laser warning labels in German and French
are included in the scope of delivery.
•
If the devices are installed in a system/casing so that the warning labels are cov‐
ered, other warning labels (not included with delivery) must be attached to the
system/casing next to the laser output aperture.
•
When operating with the trigger, the laser diode is switched on by the switch-on
signal and switched off again by the deactivation signal. When free output of mea‐
sured values is set, the laser diode is continuously on.
•
The device automatically monitors beam generation and automatically shuts down
the laser diode in the event of irregularities. If this happens, the red “Status” LED
lights up. The device stops sending measured values and an entry is added to the
error memory.