- 14 - 

 

Starting Operation

 

 

Start  Date

Configuration  item  for  when  Timer  or  Timer&Acc  trigger  mode  is  selected. 

Measuring will started at the time you selected. You do not need to configure 

this in Immediately or Acc. mode.   

START

Measuring will begin after clicking, at the time as appeared on the PC. When 

“Timer” 

or “Timer&Acc.” modes are selected, the unit will start to measure at your selected 

time automatically after click “START”. 

STOP

Measuring will be stopped after clicking, and the time on the PC will appeared. 

 

Record Data

 

 

Trig. Data

Measured acceleration data is shown. 

At Immediately or Timer mode, trig data is shown as 1. 

At Acc. or Timer&Acc. mode, trig. data shows the number of measured data of 

each frame you selected. 

Start Date   

Date & time of download shows here. 

Acc.

G

     

Selected acceleration range is shown. 

Sampling(ms)

Selected sampling rate is shown. 

Frame         

  :

Selected frame is shown. 

Folder           

Stored folder for data download is selected as you require. 

Download       

Start data download from AccStick. Data will be appeared on the screen and 

converted csv file in the selected folder. 

 

 

 

 

 

 

Содержание A01-16400

Страница 1: ...annot be responsible for any mistakes which might have occurred within its contents Please inform us if you have found obscurities or wrong descriptions skipped descriptions or the like HEAD OFFICE B4...

Страница 2: ......

Страница 3: ...andling of AccStick 3 3 2 2 Data capturing software 4 4 Operation 5 4 1 Battery 5 4 2 Body 6 4 3 Module for data connection charging of battery 6 4 4 Data capacity 7 4 5 Data file 7 5 Steps for instal...

Страница 4: ......

Страница 5: ...k in any place with heats over 60 or extremely wet These might be cause problems Prohibit Do not expose AccStick to wet conditions as damage or heating the internal battery might be occurred Prohibit...

Страница 6: ...hrough the specified software and converted to csv file 2 1 Features 2 1 1 Instrument a 3D MEMS accelerometer 16G 200G b Small size light weight c Built in Timer d Lithium polymer rechargeable battery...

Страница 7: ...his unit in any wet conditions In an environment where the instrument is exposed to the direct sunlight the functions of the electronic circuits and a lithium battery might be severely affected d The...

Страница 8: ...By using the software you are agreeing to the following articles a To use this software Windows 7 10 is necessary as an OS Install the OS onto the hard disc in advance b We are not responsible for any...

Страница 9: ...measuring condition etc Note1 After measuring turn off the main power to prevent any issues with over charging of battery Note2 Measurements should be made during the span of time in which the batter...

Страница 10: ...urn off the power after data has been downloaded to your PC successfully Also retain the power turn off condition after measuring to prevent any issues with over discharge b Connector for connection w...

Страница 11: ...data can be up to a maximum of 500 000 data points with the selected frame length set as 1 frame Refer to table below for the relationship between frame length and number of capture able data Relatio...

Страница 12: ...Stick html Execute setup exe in the folder Click install if a warning screen like the right screen appears Your installation is complete when the screen below appears USB driver install Please install...

Страница 13: ...tween the USB cable module harness and AccStick Install the specific USB driver to your PC Connect AccStick to PC with the cable module like the picture on the right Charge the battery if necessary US...

Страница 14: ...Accstick after data has been downloaded successfully Starting Conditions Select the start date when you select Timer or Timer Acc Then click START Measuring will start automatically at the selected t...

Страница 15: ...f successful Register Upload to the AccStick the measuring conditions you ve selected Letter on the icon change to blue one if successful START Start to measure on the AccStick Time at clicking is sho...

Страница 16: ...ng monitoring function Monitoring data real time can be displayed on the screen after click Monitor while connected between the AccStick and PC To stop the monitoring click Monitor again The vertical...

Страница 17: ...sec Measuring time of 1 frame obtained by Frame Sampling is shown Acc G Measurable acceleration range is selectable from 2 to 400G as you require Sampling ms Sampling for acceleration capturing is se...

Страница 18: ...ing will be stopped after clicking and the time on the PC will appeared Record Data Trig Data Measured acceleration data is shown At Immediately or Timer mode trig data is shown as 1 At Acc or Timer A...

Страница 19: ...fer to P 10 Main power of AccStick was turned off during the measuring Main power should be turned on at the switch on AccStick before starting refer to P 10 After measuring the main power was turned...

Страница 20: ...5ms at 20 Depends on temperature and measuring conditions Operational temperature range 0 to 60 Size mass 32 4 21 0 10 5 15 0g Casing Aluminium 9 1 2 Software Trigger mode 4 modes Immediately Timer Ac...

Страница 21: ...ccur please contact your sales agent or our company We will repair the unit at counter value in the following cases even if it s within the warranty period a Problems due to usage other than normal wo...

Отзывы: