background image

 

- 7 - 

<LED operation> 

Red/ blue LED flashes when connecting to PC 

Orange LED appears when the battery is charging 

   

 

4-4

 

Data capacity 

AccStick has a 4MB flash memory can store acceleration data up to 500,000 points. 

Measuring will be continuous until the selected frame length is achieved when trigger mode is 

configured as “immediately” or “Timer”. 

When in 

“Acc.”or “Timer&Acc.”mode, the captured data can be up to a maximum of 500,000 

data  points  with  the  selected  frame  length  set  as  1  frame.  Refer  to  table  below  for  the 

relationship between frame length and number of capture-able data. 

 

Relationship between Frame and Data number 

Selected Frame 

Capturable data number 

500 

1000 

1000 

500 

10000 

50 

500000 

 

4-5

 

Data file 

Measured  data  is  not  just  displayed  on  the  PC  software  but  also  converted  to  .csv  file  after 

downloading to your PC.   

The title of the .csv file will appear like ACCSTICK-YYMMDD_hhmmss. 

(YY means measurement year, MM means month, hh:mm:ss is the time information.)   

 

Содержание A01-16400

Страница 1: ...annot be responsible for any mistakes which might have occurred within its contents Please inform us if you have found obscurities or wrong descriptions skipped descriptions or the like HEAD OFFICE B4...

Страница 2: ......

Страница 3: ...andling of AccStick 3 3 2 2 Data capturing software 4 4 Operation 5 4 1 Battery 5 4 2 Body 6 4 3 Module for data connection charging of battery 6 4 4 Data capacity 7 4 5 Data file 7 5 Steps for instal...

Страница 4: ......

Страница 5: ...k in any place with heats over 60 or extremely wet These might be cause problems Prohibit Do not expose AccStick to wet conditions as damage or heating the internal battery might be occurred Prohibit...

Страница 6: ...hrough the specified software and converted to csv file 2 1 Features 2 1 1 Instrument a 3D MEMS accelerometer 16G 200G b Small size light weight c Built in Timer d Lithium polymer rechargeable battery...

Страница 7: ...his unit in any wet conditions In an environment where the instrument is exposed to the direct sunlight the functions of the electronic circuits and a lithium battery might be severely affected d The...

Страница 8: ...By using the software you are agreeing to the following articles a To use this software Windows 7 10 is necessary as an OS Install the OS onto the hard disc in advance b We are not responsible for any...

Страница 9: ...measuring condition etc Note1 After measuring turn off the main power to prevent any issues with over charging of battery Note2 Measurements should be made during the span of time in which the batter...

Страница 10: ...urn off the power after data has been downloaded to your PC successfully Also retain the power turn off condition after measuring to prevent any issues with over discharge b Connector for connection w...

Страница 11: ...data can be up to a maximum of 500 000 data points with the selected frame length set as 1 frame Refer to table below for the relationship between frame length and number of capture able data Relatio...

Страница 12: ...Stick html Execute setup exe in the folder Click install if a warning screen like the right screen appears Your installation is complete when the screen below appears USB driver install Please install...

Страница 13: ...tween the USB cable module harness and AccStick Install the specific USB driver to your PC Connect AccStick to PC with the cable module like the picture on the right Charge the battery if necessary US...

Страница 14: ...Accstick after data has been downloaded successfully Starting Conditions Select the start date when you select Timer or Timer Acc Then click START Measuring will start automatically at the selected t...

Страница 15: ...f successful Register Upload to the AccStick the measuring conditions you ve selected Letter on the icon change to blue one if successful START Start to measure on the AccStick Time at clicking is sho...

Страница 16: ...ng monitoring function Monitoring data real time can be displayed on the screen after click Monitor while connected between the AccStick and PC To stop the monitoring click Monitor again The vertical...

Страница 17: ...sec Measuring time of 1 frame obtained by Frame Sampling is shown Acc G Measurable acceleration range is selectable from 2 to 400G as you require Sampling ms Sampling for acceleration capturing is se...

Страница 18: ...ing will be stopped after clicking and the time on the PC will appeared Record Data Trig Data Measured acceleration data is shown At Immediately or Timer mode trig data is shown as 1 At Acc or Timer A...

Страница 19: ...fer to P 10 Main power of AccStick was turned off during the measuring Main power should be turned on at the switch on AccStick before starting refer to P 10 After measuring the main power was turned...

Страница 20: ...5ms at 20 Depends on temperature and measuring conditions Operational temperature range 0 to 60 Size mass 32 4 21 0 10 5 15 0g Casing Aluminium 9 1 2 Software Trigger mode 4 modes Immediately Timer Ac...

Страница 21: ...ccur please contact your sales agent or our company We will repair the unit at counter value in the following cases even if it s within the warranty period a Problems due to usage other than normal wo...