1 How to Begin
LF(R) Series Operation Manual
- 4 -
1-2 For More Precise Measurements
To be able to conduct more precise measurements, it is necessary to minimize the factors
that contribute to measurement errors. There are a great variety of such error-inducting
factors, which can be linked to machine error and performance of the balance itself, as well
as the properties and condition of samples being measured, and the measuring environment
(e.g., vibration, temperature/humidity). These factors can readily affect the results of
measurement on a balance that has high resolution readability.
This material includes some precautionary notes that the user should bear in mind to
eliminate error factors and ensure accurate measurement results.
1-2-1 Precautions on the Measuring Room
Temperature /
Humidity
•
Try to maintain constant room temperature as much as possible to prevent dew
condensation and unstable indications due to fluctuations in temperature.
•
Low relative humidity tends to induce static electricity, causing measurement error.
(Relative humidity of about 60% is considered ideal.)
Vibration /
Shaking
•
The measuring room should preferably be located on the ground floor or in the
basement. Higher floors are more susceptible to heavy vibration and shaking, which
make such locations less suitable for measurement. A room facing a railway or road
with heavy traffic should also be avoided as much as practicable.
Drafts
•
Avoid choosing a location subject to a direct draft of airflow from an air-conditioning
unit or exposed to direct sunlight, which may cause unstable reading due to abrupt
fluctuations in temperature.
•
Also avoid a room subject to a heavy flow of people, since fluctuations in drafts and
temperature are likely to occur in such a location.
Gravity
•
The gravity acting on a sample varies depending on the latitude or height of the
location where measurement is being conducted. For this reason, the same sample
may show different weight indications from one place to another.
Therefore, make it a rule to calibrate the balance every time it is relocated.
Electromagnetic
Waves
•
When the balance is located near an object that generates intense electromagnetic
waves, it may be hindered from showing accurate weight due to the effects of such
waves. Therefore, avoid placing the balance in such a location.
Содержание LF124R
Страница 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Страница 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Страница 132: ...Appendixes LF R Series Operation Manual 122 ...
Страница 133: ...LF R Series Operation Manual Appendixes 123 ...
Страница 134: ...Appendixes LF R Series Operation Manual 124 ...
Страница 135: ...LF R Series Operation Manual Appendixes 125 ...
Страница 136: ...Appendixes LF R Series Operation Manual 126 ...
Страница 137: ...LF R Series Operation Manual Appendixes 127 ...
Страница 138: ...Appendixes LF R Series Operation Manual 128 ...
Страница 139: ...LF R Series Operation Manual Appendixes 129 ...
Страница 142: ...Appendixes LF R Series Operation Manual 132 Span test with internal weight Span test with external weight ...
Страница 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Страница 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Страница 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Страница 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Страница 153: ......
Страница 154: ......