7 Troubleshooting
LF(R) Series Operation Manual
- 116 -
Problem Cause Remedy
Communication conditions
do not match those of the
external device.
referring to their operation
manuals.
Check output by:
Initialize the function
settings by referring to
Section 7-3 “Initializing” or
Function Setting List.
Then, set communication
conditions 1,200 bps, 8-bit
data, 2-bit stop bit, and no
parity by referring to the
operation manual of the
external device.
With these settings, after
the [Print] key is pressed
and once the balance has
stabilized, output is done
once. Check the output by
pressing the [Print] key.
Wrong cable connection
To connect the balance to a
common PC, a crossover
cable is required (Refer to
Section 6-2 “Connecting to
External Devices via the
RS-232C Interface”).
Obtain one yourself or
contact the retailer from
whom the balance was
purchased.
The cable is disconnected
or not properly connected.
Make sure that the cable is
connected properly.
To reset the balance to the
initial settings at the time of
purchase.
The balance can be
re-initialized (Section 7-3
“Initializing”).
Содержание LF124R
Страница 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Страница 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Страница 132: ...Appendixes LF R Series Operation Manual 122 ...
Страница 133: ...LF R Series Operation Manual Appendixes 123 ...
Страница 134: ...Appendixes LF R Series Operation Manual 124 ...
Страница 135: ...LF R Series Operation Manual Appendixes 125 ...
Страница 136: ...Appendixes LF R Series Operation Manual 126 ...
Страница 137: ...LF R Series Operation Manual Appendixes 127 ...
Страница 138: ...Appendixes LF R Series Operation Manual 128 ...
Страница 139: ...LF R Series Operation Manual Appendixes 129 ...
Страница 142: ...Appendixes LF R Series Operation Manual 132 Span test with internal weight Span test with external weight ...
Страница 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Страница 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Страница 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Страница 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Страница 153: ......
Страница 154: ......