Starting Inspection/Ending Inspection

 Instruction Manual Chapter 9 Appendix

Meaning of 

Notify

: If you discover an abnormality, notify your Shimadzu service agent.

S

tarting Inspection

! Always carry out the inspection described

below before using the system.

Before turning the power ON

Visual inspection

tel:

Responsible person:

Have unnecessary items in the vicinity of the

system been cleared away?

Clear them away.

Have previously used contrast media and drugs

been removed from the system?

Clean up.

Have you cleared away any rust and metal

fragments from the moving parts of the system

and the floor?

Clear them away.

Are any cables trapped, twisted, or rubbing

against anything?

Correct this.

Have any cable connectors been wet with water?

Clear them away.

Is there any damage to cable insulation?

Notify

Is the cart cable connected correctly?

Connect it.

Is the equipotential earth connected correctly? 

Connect it.

Is the main power cable connected correctly?

Connect it.

Inspection by touch

Do the casters, brakes, and handles of the

C-arm cart function correctly?

Notify

Do the brakes of the monitor cart function

correctly?

Notify

Is the C-arm cart’s brake applied?

Operate the brake.

Are the monitor cart’s brakes applied?

Operate the brake.

Visual/aural inspection

Is the operating noise normal?

Notify

Is any error message displayed?

(on the C-arm main panel or monitor)

Notify

After turning the power ON

E

nding Inspection

! Always check the following points before

finishing use of the system.

Before turning the power OFF

Visual inspection

Is the C-arm at its home position (fixed position)?

 

Operate the C-arm.

System operation

Is there an abnormal noise or chattering

when the C-arm is operated?

Notify

Does the C-arm lock switch function correctly?

Notify

Is the laser pointer focused correctly?

Notify

Visual inspection

Have unnecessary items in the vicinity of the

system been cleared away?

Clear them away.

Have used contrast media and drugs been

removed from the system?

Clean up.

Are any cables trapped, twisted, or rubbing

against anything?

Correct this.

Have any cable connectors been wet with

water?

Clear them away.

Is there any damage to cable insulation?

Notify

Inspection by touch

Do the casters, brakes, and handles of the

C-arm cart function correctly?

Notify

Do the brakes of the monitor cart function

correctly?

Notify

Are the monitor cart’s brakes applied? 

Operate the brake.

Is the C-arm cart’s brake applied?

Operate the brake.

After turning the power OFF

M516-E044

Содержание OPESCOPE ACTENO

Страница 1: ...art C Memory Frame Memory Loop Memory Apply power only to the C arm Conducting Examinations 1 2 Starting an Examination Selecting the APR Conducting Examinations 2 3 Fluoroscopy Adjusting Image Changi...

Страница 2: ...C arm cart s brake applied Operate the brake Are the monitor cart s brakes applied Operate the brake Visual aural inspection Is the operating noise normal Notify Is any error message displayed on the...

Страница 3: ...wn 8 1 2 4 6 7 Power switch Take the power plug out of the outlet 5 Shut down the system from the reference monitor 3 2 Main panel 3 Front of the monitor cart Back of the monitor cart C arm M516 E044...

Страница 4: ...ment C arm slide motion About the scales The zero on each scale is the C arm s home position fixed position Starting up and Shutting down the System Instruction Manual 3 6 Positioning Handle Handle Ha...

Страница 5: ...he LAN cable to the back of the monitor cart Temporary data information can be changed afterward in the Today tab or the All tab in the Study List window The patient information of images acquired in...

Страница 6: ...ic Screen APR Selection Screen Immediate Selecting the APR Immediate Select the APR 2 The basic screen is displayed and the parameters registered in the APR are loaded 1 The APR selection screen Immed...

Страница 7: ...splayed at the top right of the LCD display Resets the image brightness setting to the default value Parallel compensating filter open close Parallel compensating filter rotation Octagonal collimator...

Страница 8: ...the Dose Change the Pulse rate Changing the filter Auto Manual Switches the X ray adjustment mode The filter is changed each time is pressed Display Material Applications Filter 1 Cu0 1 mm The limbs...

Страница 9: ...r foot from the foot switch during irradiation will end X ray irradiation partway through and stop radiography Depress the right foot switch 3 During X ray irradiation the radiography image is display...

Страница 10: ...mators Beam hardening filter Image rotation and inversion I I field size Storing Calling Up Basic screen 1 Press Save You can save the currently displayed images If any condition is changed from the s...

Отзывы: