Shimadzu OPESCOPE ACTENO Скачать руководство пользователя страница 6

Conducting Examinations 1

 

Instruction Manual 2.2.2 LCD Display, 3.8 Selecting the APR

C

on

du

cti

ng

 E

xa

m

in

at

io

ns

 1

• Basic screen (Simple mode)

• Basic screen (Expert mode)

M516-E044

Startup Screen

There are two modes of operation display in the basic screen, Simple Mode and Expert Mode.

Basic Screen

APR Selection Screen (Immediate)

Selecting the APR (Immediate)

Select the APR.

2

The basic screen is displayed and the 

parameters registered in the APR are loaded.

1

The APR selection screen (Immediate) is 

displayed.

Turn 

 (Releasing the restriction on X-ray 

exposure) ON (lit).

Set [Dose], [Pulse rate] and [Loop].

When in simple mode, displays the 

fluoroscopy mode selection screen.

Displays the APR selection screen and select 
APR.

Saves the image.

Used for film radiography.
(Only Expert mode)

Содержание OPESCOPE ACTENO

Страница 1: ...art C Memory Frame Memory Loop Memory Apply power only to the C arm Conducting Examinations 1 2 Starting an Examination Selecting the APR Conducting Examinations 2 3 Fluoroscopy Adjusting Image Changi...

Страница 2: ...C arm cart s brake applied Operate the brake Are the monitor cart s brakes applied Operate the brake Visual aural inspection Is the operating noise normal Notify Is any error message displayed on the...

Страница 3: ...wn 8 1 2 4 6 7 Power switch Take the power plug out of the outlet 5 Shut down the system from the reference monitor 3 2 Main panel 3 Front of the monitor cart Back of the monitor cart C arm M516 E044...

Страница 4: ...ment C arm slide motion About the scales The zero on each scale is the C arm s home position fixed position Starting up and Shutting down the System Instruction Manual 3 6 Positioning Handle Handle Ha...

Страница 5: ...he LAN cable to the back of the monitor cart Temporary data information can be changed afterward in the Today tab or the All tab in the Study List window The patient information of images acquired in...

Страница 6: ...ic Screen APR Selection Screen Immediate Selecting the APR Immediate Select the APR 2 The basic screen is displayed and the parameters registered in the APR are loaded 1 The APR selection screen Immed...

Страница 7: ...splayed at the top right of the LCD display Resets the image brightness setting to the default value Parallel compensating filter open close Parallel compensating filter rotation Octagonal collimator...

Страница 8: ...the Dose Change the Pulse rate Changing the filter Auto Manual Switches the X ray adjustment mode The filter is changed each time is pressed Display Material Applications Filter 1 Cu0 1 mm The limbs...

Страница 9: ...r foot from the foot switch during irradiation will end X ray irradiation partway through and stop radiography Depress the right foot switch 3 During X ray irradiation the radiography image is display...

Страница 10: ...mators Beam hardening filter Image rotation and inversion I I field size Storing Calling Up Basic screen 1 Press Save You can save the currently displayed images If any condition is changed from the s...