background image

Conducting Examinations 2

M516-E044

 Instruction Manual 3.9 Fluoroscopy

C

on

du

cti

ng

 E

xa

m

in

at

io

ns

 2

Fluoroscopy

Adjusting Image

L

F

Foot switch

If the fluoroscopy timer beep sounds

If the fluoroscopy timer preset time is exceeded, 

the beep sounds.

The alarm can be stopped by pressing 

fluoroscopy timer beep sound reset switch.

Adjusting Contrast and Brightness

Collimator and Parallel Compensating Filter Operation

Image Rotation and Inversion

Automatically adjusts the contrast and 

brightness of the image in 7 stages (-2 

to +4).

The set image brightness is displayed 

at the top right of the LCD display.

Resets the image brightness setting to the default value.

Parallel compensating filter open/close

Parallel compensating filter rotation

Octagonal collimator open/close

  The image is inverted in the up/down direction.

  The image is inverted in the left/right direction.

Press 

 and then

press 

 to rotate 

the image at high speed.

The image rotates.

Before moving away from the system for 

a long time,

Turn 

 (Releasing the restriction on X-ray 

exposure) OFF (lamp off) in order to 

prevent 

malfunction.

Memo

1

During radiation, the irradiation 

indicator (on the monitor cart) 

is lit. The fluoroscopic image is 

displayed on the monitor.

Press X-ray exposure 

(fluoroscopy) switch or 

depress the left foot 

switch.

Irradiation indicator

Lit

Содержание OPESCOPE ACTENO

Страница 1: ...art C Memory Frame Memory Loop Memory Apply power only to the C arm Conducting Examinations 1 2 Starting an Examination Selecting the APR Conducting Examinations 2 3 Fluoroscopy Adjusting Image Changi...

Страница 2: ...C arm cart s brake applied Operate the brake Are the monitor cart s brakes applied Operate the brake Visual aural inspection Is the operating noise normal Notify Is any error message displayed on the...

Страница 3: ...wn 8 1 2 4 6 7 Power switch Take the power plug out of the outlet 5 Shut down the system from the reference monitor 3 2 Main panel 3 Front of the monitor cart Back of the monitor cart C arm M516 E044...

Страница 4: ...ment C arm slide motion About the scales The zero on each scale is the C arm s home position fixed position Starting up and Shutting down the System Instruction Manual 3 6 Positioning Handle Handle Ha...

Страница 5: ...he LAN cable to the back of the monitor cart Temporary data information can be changed afterward in the Today tab or the All tab in the Study List window The patient information of images acquired in...

Страница 6: ...ic Screen APR Selection Screen Immediate Selecting the APR Immediate Select the APR 2 The basic screen is displayed and the parameters registered in the APR are loaded 1 The APR selection screen Immed...

Страница 7: ...splayed at the top right of the LCD display Resets the image brightness setting to the default value Parallel compensating filter open close Parallel compensating filter rotation Octagonal collimator...

Страница 8: ...the Dose Change the Pulse rate Changing the filter Auto Manual Switches the X ray adjustment mode The filter is changed each time is pressed Display Material Applications Filter 1 Cu0 1 mm The limbs...

Страница 9: ...r foot from the foot switch during irradiation will end X ray irradiation partway through and stop radiography Depress the right foot switch 3 During X ray irradiation the radiography image is display...

Страница 10: ...mators Beam hardening filter Image rotation and inversion I I field size Storing Calling Up Basic screen 1 Press Save You can save the currently displayed images If any condition is changed from the s...

Отзывы: