41
O
pe
ra
tin
g Cam
er
a
High-Speed Video Camera
HPV-X2
■
Selecting Exposure Time for Recording in [REC]
– [EXPOSE]
Exposure time for recording is selected in the [REC] – [EXPOSE] field.
The default recording exposure time setting after initial startup is selectable from 200 ns,
500 ns, 1,000 ns, 2,000 ns, 5,000 ns, 10,000 ns, 20,000 ns, 50,000 ns, 100,000 ns,
1,000,000 ns, 2,000,000 ns, 5,000,000 ns, or 10,000,000 ns.
In addition, clicking the
icon displays a window for editing the recording exposure
time selection list (Fig. 5-12). This list of selectable recording exposure time settings can
be edited.
Fig. 5-12 Window for Editing the Recording Exposure Time Selection List
Operations in the Window for Editing the Recording Exposure Time
Selection List
Procedure for adding exposure time settings to the exposure time selection list
1.
Display the window for editing the exposure time selection list (Fig. 5-12).
2.
In the field to the right of the [<< Add] button, enter the exposure time setting to
be added to the list.
3.
Click [<< Add] to add the exposure time setting entered in step 2 to the [LIST]
field.
4.
Click [OK].
Save the settings and close the window. Once the settings are saved, the added
exposure time settings are selectable in [REC] – [EXPOSE] in the [REC] window.
To not save the settings, click [CANCEL].
Procedure for deleting exposure time settings from the exposure time selection
list
1.
Display the window for editing the exposure time selection list (Fig. 5-12).
2.
Select the exposure time setting to delete from the [LIST] field.
3.
Click [Delete >>] to delete the exposure time setting selected in step 2.
4.
Click [OK].
Save the settings and close the window. Once the settings are saved, the deleted
exposure time settings are no longer selectable in [REC] – [EXPOSE] in the [REC]
window.