44
Error Messages Related to FPD
When "FPD ERROR" is displayed at upper left of the system monitor, errors in the following list are
also displayed.
[HV/DC Power]:Emergency
shutdown
[HV Power Supply]:Over Current
[DC Power Supply]:Overload
[Sensor]:Temperature error
Stop study immediately
and contact Shimadzu
service representative.
[HV Power Supply]:Temperature
Over
[Cooling Unit]:Water flow error
Error of HV power
temperature of FPD.
Error of HV power electric
current output of FPD. HV
and DC power of FPD shut
down immediately.
Fluoroscopy and radiography
become disabled.
HV and DC power of FPD
shut down immediately.
Fluoroscopy and radiography
become disabled.
Error of FPD cooling unit
water flow. HV and DC power
of FPD shut down after 30
minutes.
DC power of FPD is
overloaded.
Error of sensor chassis
temperature. HV and DC
power of FPD shut down after
30 minutes.
Action
Message
Description