33
4
Fluoroscopy/
Radiography
Left Ventricle Analysis (LV)
■
Perform LV analysis with a projected image of RAO 30 degrees (or images of RAO 30 degree and
LAO 60 degree).
LV Tool Bar
1
Click
and select
LV Tool bar will appear.
2
Perform calibration
To perform calibration, display the image of
reference object (such as a ball) previously acquired
with the same magnification as the analysis object.
The smaller reference object (the smaller the ball),
the greater the analysis error.
1
Click
from [Calibration] menu.
2
Input the ball diameter (mm) which is on the
image.
3
Outline the ball by clicking 3 points.
Calibration factor (mm/pixel) is displayed.