15
2
System
Startup and
Shutdown
C-arm
1
Remove any object from FPD
field of view
1
Make sure that [PARK] button on the local
console is lit.
If [PARK] button is not lit, press until it is to
park the C-arm.
2
Move the Compensation Filters to the
outside.
The Compensation Filters move away from
the field of view.
3
Move the tabletop out of the FPD field of
view.
Calibration data cannot be accurately acquired if there is any
object between the X-ray Tube and FPD.
NOTE
2
Remove Grid
Tabletop
X-ray Tube
FPD
Grid
FPD