-15-
9. % SETTING
This balance serves percent (%) display by setting the reference sample to 100%.
Step (1)
Register the % unit. (Refer to “8. Registration, cancel, and change of
unit”.
When % unit has been already registered,
it is not necessary to register again.
(2)
Place the tare on the pan and press the
TARE key.
(3)
Load the reference sample.
(4)
Continuously press the UNIT key to
display “ ”.
(5)
After the stability mark is lit, press the
TARE key.
(6)
“ ” is displayed for several seconds and the balance enters
the % unit display.
Minimum displayed value changes as follows depending on reference
sample weight (REF.)
Referense sample weight (REF.) is count value which minimum
displayed value rearranges one count.
If the % conversion is not possible, “ ” is displayed for
several seconds and the balance returns to weight display.
Type H
REF. < Minimum displayed value × 100
% conversion impossible
Minimum displayed value × 100
< REF. < Minimum displayed value × 1000
100%
Minimum displayed value × 1000
< REF. < Minimum displayed value × 10000
100.0%
Minimum displayed value × 10000
< REF. < Minimum displayed value × 100000
100.00%
Minimum displayed value × 100000 < REF.
100.000%
Type S
REF. < Minimum displayed value × 100
% conversion impossible
Minimum displayed value × 100
< REF. < Minimum displayed value × 1000
100%
Minimum displayed value × 1000
< REF. < Minimum displayed value × 10000
100.0%
Minimum displayed value × 10000 < REF.
100.00%
Содержание BL2200H
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 10: ... 2 Display and keypad Display Keypad ...
Страница 36: ......