-4-
(3)
Place the pan supporter on the pan shaft
and then the pan on the pan supporter.
(4)
Plug the AC adapter into the outlet.
The balance shows
after
self-checking.
(5)
Press the POWER/BRK key. All
displays light for one minute. Then
the display automatically shows "zero"
and the balance enters measurement
ready state.
(6)
Press the POWER/BRK key again.
The standby mark lights up and the
balance enters standby state.
(7)
Warm up the balance.
(8)
Calibrate the sensitivity. Refer to “7. SENSITIVITY
CALIBRATION”.
(9)
Check the performance. Refer to “11. PERFORMANCE
CHECKS”
3.
WARM-UP
•
Apply power in advance for one hour
or more. This will allow you to
immediately make an accurate
measurement.
•
Even if the balance is not used, keep
the standby mark lit (warm-up state)
by pressing the POWER/BRK key without disconnecting
the AC adapter.
•
If the balance is not used for one month or more, disconnect
the AC adapter.
Содержание BL2200H
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 10: ... 2 Display and keypad Display Keypad ...
Страница 36: ......