13. Screw the lid securely onto the sensor.
Re-install the Sensor
1. Carefully slide the sensor into the analyzer a few inches.
2. Re-connect the quick disconnect at the top of the flowmeter. (Skip this
step if the analyzer is equipped with the stainless steel outlet tubing
option.)
3. Carefully slide the sensor the rest of the way into the analyzer.
4. If the analyzer is equipped with the stainless steel outlet tubing option,
replace the nut from the outlet bulkhead fitting.
5. Replace the bulkhead nut to the inlet gas connection on the rear of the
analyzer.
6. Secure the two knurled thumb screws and verify that the sensor cannot
move about in the analyzer.
7. Re-connect the electrical cable that runs to the sensor.
Draining Sensors mounted on remote brackets and in R4 and R7
enclosures
1. Make up the supplied compression fitting, ferrule and ¼ inch PVC tube to
the drain valve on the side of the sensor.
2. Place the end of the PVC tube in a suitable container with a minimum
capacity of .75 liters.
3. Remove the cap on the top of the sensor.
4. Open the drain valve on the side of the sensor.
5. Allow sufficient time for the electrolyte to drain completely.
6. Flush the sensor with .5 liters of distilled or de-ionized water.
7. Close the drain valve and remove the tube.
Содержание DF-130E
Страница 1: ...Oxygen Analyzer DF 150E OPERATOR MANUAL...
Страница 10: ......
Страница 13: ...Specifications DF 150E 9 Figure 1 DF Series Oxygen Analyzer...
Страница 14: ......
Страница 44: ......
Страница 56: ......
Страница 84: ......
Страница 87: ...Index DF 150E 83 U Upsets Oxygen Shock 44 V Voltage 15 46 checking 15 W Warranty 79...