Options DF-150E
21
The alarm relays are configured for “Fail Safe” operation. The relays will go to
an Alarm Condition when the analyzer is turned off or when power fails.
7.2.2.1 Non-CE Qualified Analyzers
Each alarm has a SPDT relay rated at 125/250 VAC at 5 Amps or 30 VDC at
5 Amps resistive load.
7.2.2.2 CE Qualified Analyzers
Each alarm has an SPDT relay rated at 30 VDC at 5 Amps. Exceeding this
rating may cause damage to the analyzer. Not to be used for switching AC
voltage.
7.3 Pressure Regulators
The Standard Regulator is rated for 3000 psig inlet pressure.
The Absolute Regulator is rated to 1000 psig inlet pressure.
All brackets and plumbing are provided with this option.
Figure 9: Regulator Installation