![SEFRAM 7334 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/sefram/7334/7334_user-manual_1233374040.webp)
19
(8) Capacitance
Range
Resolution
Accuracy
1.000
μ
F 0.001
μ
F
± (1.2% + 2d)
10.00
μ
F 0.01
μ
F
100.0
μ
F 0.1
μ
F
1.000mF 0.001mF
10.00mF 0.01mF
Input Protection :
1000V
DC
or 1000V
AC
rms
Max. Measuring Time :
0.7 sec for 1nF~1mF
3 sec for 1mF~10mF
(9) Frequency Counter
Range
Resolution
Accuracy
100.00 Hz
0.01 Hz
± (0.1% + 2d)
1000.0 Hz
0.1 Hz
10.000 KHz
0.001 KHz
100.00 KHz
0.01 KHz
Input Protection :
1000V
DC
or 1000V
AC
rms
Min. Frequency :
1Hz
Sensitivity :
>
5.0V
p-p
(for ACV 1HZ ~ 10kHz).
>
10V
p-p
(for ACV 10kHZ ~ 100kHz).
>
2mA
p-p
(for ACmA).
>
0.2A
p-p
(for ACA).
(10) Temperature (For 7335 only)
Range
Resolution
Accuracy*
-40.0°C ~ 400°C
0.1°C
± (1.0% + 10d)
-40.0°F ~ 752°F
0.1°F
± (1.0% + 18d)
Input Protection
:
1000V
DC
or 1000V
AC
rms
* Does not include accuracy of the thermocouple probe.
* Accuracy specification assumes surrounding temperature
stable to ± 1°C. For surrounding temperature changes of
± 5°C, rated accuracy applies after 2 hours.