16
EMC :
EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-PRF- 28800F
(5 ~ 55 Hz, 3g max.)
Drop Protection :
4 feet drop to hardwood on concrete floor.
Indoor Use.
Electrical Specifications
Accuracy is ±(% r number of digits) at 23°C ± 5°C
< 80%RH.
(1) Voltage
Range
Resolution
For 7334
Accuracy
60.00mV
DC
0.01mV
600.0mV
DC
0.1mV
± (0.1% + 2d)
6.000V
DC
0.001V
± (0.09% + 2d)
60.00V
DC
0.01V
± (0.09% + 2d)
600.0V
DC
0.1V
± (0.09% + 2d)
1000V
DC
1V
± (0.09% + 2d)
60.0mV
AC
0.01mV
600.0mV
AC
0.1mV
± (1.5% + 5d)
6.000V
AC
0.001V
± (1.0% + 3d)
60.00V
AC
0.01V
± (1.0% + 3d)
600.0V
AC
0.1V
± (1.0% + 3d)
1000V
AC
1V
± (1.0% + 3d)
For 7335
Accuracy
± (0.08% + 10d)
± (0.08% + 2d)
± (0.08% + 2d)
± (0.08% + 2d)
± (0.08% + 2d)
± (0.08% + 2d)
± (1.20% + 5d)
± (1.20% + 5d)
± (0.80% + 5d)
± (0.80% + 5d)
± (0.80% + 5d)
± (0.80% + 5d)
Input Protection
:
1000V
DC
or 1000V
AC
rms
AC Frequency Response :
50Hz ~ 1KHz
Input Impedance :
10M
Ω
, <100pF
AC Conversion Type
:
AC conversions are ac-coupled, true rms responding, calibrated
to the sine wave input.
For non-sine wave add the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.