17
(2) Auto-V
Range
Resolution
For 7334
Accuracy
For 7335
Accuracy
600.0V
DC
0.1V
± (1.0% + 3d)
± (0.80% + 3d)
1000V
DC
1V
600.0V
AC
0.1V
1000V
AC
1V
Input Protection :
1000V
DC
or 1000V
AC
rms
AC Frequency Response
:
50Hz ~ 1KHz
Input Impedance :
Approx. 3k
Ω
.
AC Conversion Type :
Conversion type and additional specification are same as
voltage function.
(3) Current
Range
Resolution
For 7334
Accuracy
For 7335
Accuracy
60.00mA
DC
0.01mA
± (1.0% + 3d)
± (0.8% + 3d)
600.0mA
DC
0.1mA
± (1.0% + 3d)
± (0.8% + 3d)
6.000A
DC
0.001A
± (1.0% + 3d)
± (0.8% + 3d)
10.00A
DC
0.01A
± (1.0% + 3d)
± (0.8% + 3d)
60.00mA
AC
0.01mA
± (1.5% + 3d)
± (1.2% + 3d)
600.0mA
AC
0.1mA
± (1.5% + 3d)
± (1.2% + 3d)
6.000A
AC
0.001A
± (1.5% + 3d)
± (1.2% + 3d)
10.00A
AC
0.01A
± (1.5% + 3d)
± (1.2% + 3d)
Input Protection
:
Equipped with High Energy Fuse.
440mA, 1000V AC/DC 10KA Fuse for mA input .
11A, 1000V AC/DC 20KA Fuse for A input.
Max. Measuring Time
:
3 minutes for A input, 10 minutes for mA input.
Rest time 20 minutes min.
AC Conversion Type
:
Conversion type and additional specification are same as volt-
age function.
AC Frequency Response
:
50Hz ~ 1KHz