![SEFRAM 7334 Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/sefram/7334/7334_user-manual_1233374039.webp)
18
(4) Peak Hold (For 7335 only) :
Specified accuracy ±150d.
(5) Resistance
Range
Resolution
Accuracy
600.0
Ω
0.1
Ω
± (0.8% + 5d)
6.000K
Ω
0.001K
Ω
± (0.8% + 2d)
60.00K
Ω
600.0K
Ω
6.000M
Ω
40.00M
Ω
* 0.01M
Ω
± (1.0% + 5d)
0.01K
Ω
0.1K
Ω
0.001M
Ω
Input Protection :
1000V
DC
or 1000V
AC
rms
Max. Open Circuit Voltage :
Approx. 2.5V for 600
Ω
& 6k
Ω
Range.
Approx. 0.6V for others.
Max. Test Current
:
Approx. 0.1mA.
* There is a little rolling less than ±50 digits when measuring
>
10.00M
Ω
.
(6) Continuity Check
Range
Resolution
Accuracy
600.0
Ω
0.1
Ω
± (0.8% + 5d)
Input Protection :
1000V
DC
or 1000V
AC
rms
Max. Open Circuit Voltage :
Approx. 2.5V
Max. Test Current
:
Approx. 0.1mA
Continuity Threshold
:
<
30
Ω
Beep On.
>
100
Ω
Beep OFF.
Continuity Indicator
:
2.7KHz Tone Buzzer
(7) Diode Test
Range
Resolution
Accuracy
2.000V
1mV
± (1.5% + 2d)
Input Protection :
1000V
DC
or 1000V
AC
rms
Max. Open Circuit Voltage :
Approx. 2.5V
Max.Test Current :
Approx. 0.4mA