15
Specifications
General Specifications
Maximum voltage applied to any terminal :
1000 V
ac
rms or 1000 V
dc
. rms
Display :
6000 counts.
Polarity Indication :
Automatic, positive implied, negative indicated.
Overrange Indication :
OL
Batteries Life :
150hours ALKALINE Battery (No Backlight)
Low Batteries Indication :
Voltage drops below operating voltage, will flash.
Power Requirement :
9V battery
Auto Power Off :
20 minutes.
Operating Temperature :
-10 ~10°C
10°C ~ 30°C (
≦
80% RH),
30°C ~ 40°C (
≦
75% RH),
40°C ~ 50°C (
≦
45%RH)
Storage Temperature :
-20°C to 60°C , 0 to 80% R.H. (batteries not fitted)
Temperature Coefficient :
0.15 x (Spec.Accy) / °C, < 18°C or > 28°C .
Measure :
Samples 3 times per second .
Altitude :
6561.7 ft (2000m)
Safety :
Complies with EN61010-1, UL61010-1,IEC 61010-1,
CAT.IV. 600V, CAT.
Ⅲ
. 1000V
CAT Application field
Ⅰ
The circuits not connected to mains.
Ⅱ
The circuits directly connected to Low-voltage installa-
tion.
Ⅲ
The building installation.
Ⅳ
The source of the Low-voltage installation.
Compliance to EN 61557 :
IEC61557-1, IEC61557-2,
IEC61557-4, IEC61557-10
Weight :
460g (including battery)
Dimensions (W x H x D) :
94mm x190mm x 48mm with holster.
Accessories :
Battery (installed), Test leads and user manual.
(The probe assembly provided with the product are for use with
this product)
Power Requirements :
9V battery.
Pollution degree :
2