background image

 

10 

 Use 

PolyLink kit 

CAUTION: 

The  Bluetooth  Low  Energy  wireless  connection  must  be  enabled  for  acquisition  of  signals  from  the 
effort/position kit. 

 

 

 

Views of the display: 

 

Tap the touch 

key 

 

 

 

 

Communication  with  the  PolyLink  effort/position 
kit 
The  S.Box  device  searches  the  PolyLink 
effort/position  kit  and  the  BLE  display  flashes 
alternately  with  six  dashes  as  long  as  no  kit  is 
paired. 
When the kit is detected, the BLE display flashes 
alternately  with  the  last  six  alphanumeric 
characters of the Bluetooth address indicated on 

the  rear  side  of  the  kit, 

  as  shown  in 

the example. 
If  the  6  characters  displayed  do  not  match  the 
PolyLink effort / position kit used, a new search is 

possible

 

by tapping the key 

 or 

Status menu 

This menu is available from the  firmware version 1.7 of the S.Box device. It is  accessible when the 
S.Box  device  is  in  operation  and  when  a  PolyLink  system  or  a  pulse  oximeter  WristOx2  has  been 
installed and paired. The information can be viewed successively

 

but cannot be modified.  

To access it, tap the 

 touch button when the S.Box device is on.  

 

❶ 

Tap the touch 

button 

 

when the device 

is in operation. 

 
 
 

 

 

 

SD card status  
 
 
Possible displays: SD: OK and SD: _ _ _. 
 
If SD: _ _ _.is displayed, check that the SD card is 
present and properly inserted in the memory card 
slot.  
 

 

❷ 

Tap the touch 

button 

 

 
 
 
 

 

 

 

 
Status of the pulse oximeter 
 
Possible displays: SPO: OK and SPO: _ _ _. 
 
If SPO: _ _ _ is displayed, check that the oximeter 
sensor  has  been  properly  installed  and  the 
communication with the oximeter is correct. 

 

 

 

Содержание PolyLink

Страница 1: ......

Страница 2: ...chnical locations SEFAM 144 AV CHARLES DE GAULLE 92200 NEUILLY SUR SEINE FRANCE SEFAM 10 ALLEE PELLETIER DOISY 54600 VILLERS LES NANCY FRANCE TEL 33 0 3 83 44 85 00 www Sefam medical com Technical assistance E mail technicalservice sefam medical com ...

Страница 3: ...9 Communication with the PolyLink kit 9 Status menu 10 Real time acquisition 12 Stopping treatment 12 Transport of the PolyLink kit 12 Removing the PolyLink kit 13 Cleaning and Caring 13 Trouble shooting 14 Technical characteristics 15 Performance of the PolyLink kit 15 Radio specifications 15 Conditions of use 15 Transport and storing conditions 15 Electrical characteristics 15 Physical features ...

Страница 4: ...ny critical medical device such as a pacemaker contact your physician and the manufacturer of this device for more precautions Do not use the PolyLink kit in locations containing flammable or explosive materials CAUTION In this manual it means that there is the possibility of material damage to this unit or any other Contact your Health Professional if you suspect that this accessory or one of its...

Страница 5: ...is equipment and the other devices should be observed to verify that they are operating normally WARNING Use of accessories transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation WARNING RF portable communication d...

Страница 6: ...the internal battery indicates the positions On right and Off left of the switch on the PolyLink effort position kit support in which the PolyLink effort position kit must be inserted It is used to position and maintain the effort sensor on the thorax belt and to store the surplus cables sliding door depending on its position you can either switch on switch off or charge the PolyLink effort positi...

Страница 7: ...n Off of the switch on PolyLink effort position kit Symbol indicating the charge state of the internal battery of the PolyLink effort position kit USB connection Class II device Non ionizing radiation symbol Manufacturer Manufacturing date Reference of the PolyLink module for S Box Serial number Device in accordance with the requirements of the European directive 93 42 EC on medical devices Device...

Страница 8: ...quipped with a humidifier reinsert the heater plate on the rails in the side part of the device and push it against the device Re lock the heater plate by pushing the latch to the left Install the base of the humidifier chamber on the heater plate and push the chamber against the device until you hear a click 5 Plug the S Box device into the power supply or electrical power grid Implementation of ...

Страница 9: ...e patient s body position to the PolyLink module for S Box Electromagnetic disturbances may temporarily interrupt the transmission of data without danger for the user The internal battery for the PolyLink effort position kit is recharged by the USB charger In this configuration electromagnetic disturbances may temporarily interrupt the charge of the battery without danger for the user Communicatio...

Страница 10: ...it used a new search is possible by tapping the key or Status menu This menu is available from the firmware version 1 7 of the S Box device It is accessible when the S Box device is in operation and when a PolyLink system or a pulse oximeter WristOx2 has been installed and paired The information can be viewed successively but cannot be modified To access it tap the touch button when the S Box devi...

Страница 11: ...ing on the importance of the thorax movements If THO _ _ _ is displayed check that PolyLink effort position kit is turned on that the thoracic effort sensor is connected and that its tension is correct Tap the touch button Status of the body position sensor Possible displays POS lying on back POS lying on stomach POS lying on right side POS lying on left side POS standing POS sensor inverted POS _...

Страница 12: ... present CAUTION Make sure that the SD card is inserted in the S Box device for the real time acquisition Check that the symbol is present in the center of the display showing that the BLE wireless communication is enabled Do not unplug the S Box device Once the S Box comes back to standby the polygraphic control is complete the SD memory card can be removed from the S Box device The data saved on...

Страница 13: ...er chamber on the heater plate and push the chamber against the device until you hear a click 5 Unplug the effort sensors from the PolyLink effort position kit 6 Refer to the user manuals to remove the effort sensors thorax and abdomen and their belts Cleaning and Caring As the PolyLink module is an accessory designed to be installed inside the S Box device it does not require any special maintena...

Страница 14: ...em if necessary If the problem persists contact your Health Professional The LED of the effort position kit is off The extinction of the LED is automatic when the effort position kit has been working for 1 minute If the LED never lights up contact your Health Professional The symbol is not present in the center of the display when the S Box device is working The BLE Bluetooth Low Energy communicat...

Страница 15: ... 0 2 500 m Transport and storing conditions Pressure range 700 hPa to 1060 hPa Temperature 25 C to 70 C Relative humidity between 5 and 93 without condensation Electrical characteristics PolyLink module for S Box Input voltage 5 Vdc 5 powered by the electrical connector of the S Box device Maximum power consumption 45 mW 9 mA PolyLink effort position kit Input voltage 5 Vdc 5 Maximum power consump...

Страница 16: ... the following directives and standards Directive 93 42 EEC of the Council concerning medical devices modified by the European Directive 2007 47 EC Directive 2014 53 EU RED of the European Parliament and of the Council on the harmonization of the laws of the Member States relating to the making available on the market of Radio Equipment Directive 2006 66 EC of the European Parliament and of the Co...

Страница 17: ... energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations Flicker emissions IEC 61000 3 3 Complies Equipment is tested in HOME HEALTHCARE ENVIRONMENT more severe than the PROFESSIONAL HEALTHCARE FACILITY ENVIRONMENT It can be connected to PU...

Страница 18: ...18 Technical characteristics PolyLink kit ...

Страница 19: ......

Страница 20: ...r Manufacturing plant SEFAM 144 AV CHARLES DE GAULLE 92200 NEUILLY SUR SEINE FRANCE SEFAM 10 ALLEE PELLETIER DOISY 54600 VILLERS LES NANCY FRANCE M 167DFU00 10 Version 2 2018 12 Contact of your Health Professional ...

Отзывы: