![SEFAM PolyLink Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/sefam/polylink/polylink_procedure-for-installation-and-usage_1233311013.webp)
PolyLink kit
Removal of the PolyLink kit
13
Removal of the PolyLink kit
1.
Unplug the S.Box device from the power supply or the electrical power grid.
2.
Remove the side cover or the humidifier from the machine:
Press the button for unlocking the cover or the humidifier chamber and pull simultaneously
the cover or the chamber towards you by using the corresponding integrated handle.
In addition, if the S.Box is equipped with a humidifier, unlock the heater plate by pushing the
latch to the right and remove it.
3.
Lift the lever of
the PolyLink module and pull the module out of the accessory location of the
device.
4.
Re-place the side cover or the humidifier into the device:
Reinsert the cover on the rails in the side part of the device and push it against the device until
you hear a "click".
If the S.Box is equipped with a humidifier, reinsert the heater plate on the rails in the side part
of the device and push it against the device. Re-lock the heater plate by pushing the latch to
the left. Place the base of the humidifier chamber on the heater plate and push the chamber
against the device until you hear a "click".
5.
Unplug the effort sensors from the
PolyLink
effort/position kit.
6.
Refer to the user manuals to remove the effort sensors (thorax and abdomen) and their belts.
Cleaning and Caring
As the PolyLink module is an accessory designed to be installed inside the S.Box device, it does not
require any special maintenance. However, if necessary, follow the procedure hereafter.
WARNING:
Unplug the S.Box device and remove the
PolyLink
module from the device before cleaning it (see paragraph
CAUTION:
To clean the
PolyLink
module, use only materials reserved for this purpose: do not use aggressive
detergent, scratching sponge or hard bristle brush.
Do not let any water enter the
PolyLink
module.
Clean the outside of the
PolyLink
module with a damp cloth sprinkled with a little water and a drop
of mild detergent.
Let it dry completely before reinstalling the
PolyLink
module into the S.Box device.
The effort/position kit can be cleaned in the same way as the PolyLink module, after disconnection of
the effort sensors.
For cleaning and caring the respiratory effort sensors and the belts, refer to their respective user
manuals.